These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Sensitive determination of a new antiarrhythmic agent, trecetilide, in plasma by high-performance liquid chromatography with fluorescence detection.
    Author: Buist SC, Hsu CY, Walters RR.
    Journal: J Chromatogr A; 1998 Dec 18; 828(1-2):259-65. PubMed ID: 9916311.
    Abstract:
    Two high-performance liquid chromatography (HPLC) methods were developed for the determination of trecetilide in plasma samples. Differing only in the addition of a derivatization step and different detection wavelengths, the two methods encompassed a wide concentration range. In both methods, plasma samples (0.1 ml) with added internal standard were applied to solid-phase extraction discs containing a non-polar/strong cation mixed-phase, washed and eluted with an acetone-acetonitrile triethylamine mixture. The eluate was evaporated to dryness, and either reconstituted and directly injected onto an HPLC column or first derivatized with 1-naphthyl isocyanate before HPLC analysis. In both methods, the separation was performed isocratically on a cyano analytical column utilizing a mobile phase composed of acetonitrile-pH 7.9 phosphate buffer (70:30, v/v). The column effluent was monitored by fluorescence detection at 290/345 nm (with derivatization) or 235/320 nm (without derivatization). The limits of detection and quantitation of the assay were 0.57 and 1.9 ng/ml, respectively, when derivatization was used, or 4.3 and 14 ng/ml, respectively, without derivatization.
    [Abstract] [Full Text] [Related] [New Search]