These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The validity of the approximation method "area/perimeter" as applied to 8 and 25-MV photon beams.
    Author: Osman GE, Tsung P.
    Journal: Radiology; 1978 Oct; 129(1):233-4. PubMed ID: 99763.
    Abstract:
    The approximation method of area/perimeter for different portal sizes used in practice for finding the equivalent squares of a rectangular field has been tested on a 35-MV Clinac Linear Accelerator for the purpose of treatment planning. Data were obtained for 8 and 25-MV x-rays at depths of maximum build-up regions, and at 10 cm depth below the surface of a suitable water phantom. Maximum variation of dose calculation from the above method is less than 1.6% for routine clinical usage while less than 3.2% for extreme rectangular field sizes. Dose variation improves as the depth in phantom increases.
    [Abstract] [Full Text] [Related] [New Search]